Takumi Yanagawa, Ryosuke Abe, and Yoshio Hayasaki
Three-dimensional mapping of fluorescent nanoparticles was performed by using incoherent digital holography. The positions of the nanoparticles were quantitatively determined by using Gaussian fitting of the axial- and lateral-diffraction distributions through position calibration from the observation space to the sample space. It was found that the axial magnification was constant whereas the lateral magnification linearly depended on the axial position of the fluorescent nanoparticles. The mapping of multiple fluorescent nanoparticles fixed in gelatin and a single fluorescent nanoparticle manipulated with optical tweezers in water were demonstrated.